FOXBORO

Foxboro P0400TH-F Control Module

INTELLIGENT TRANSMITTER FEATURES AT

AN ECONOMICAL PRICE

When you want the flexibility and performance of a

configurable, intelligent transmitter, but you don’t

need a digital output signal, these transmitters

provide exceptional benefits at a very affordable

price:

 Liquid Crystal Display (LCD) Digital Indicator

with On-Board Pushbuttons

 Pushbutton Configuration and Calibration:– Zero and Span Settings– Adjustable Damping– Forward or Reverse Output– Failsafe Output; upscale or downscale– Reranging without applying pressure

GENERAL SPECIFICATIONS COMMON TO ALL TRANSMITTERS

I/A Series Pressure Transmitter Family

These transmitters are part of a complete family of

d/p Cell®, gauge, absolute, multirange,

multivariable, and premium performance

transmitters, as well as transmitters with remote or

direct connect pressure seals, all using field-proven

silicon strain gauge sensors and common topworks.

Select the electronics module you need to provide

just the right level of intelligence for your application

and budget. If your needs change, the modular

design allows easy migration to other protocols –

including FoxCom™, HART, FOUNDATION fieldbus,

and analog 4 to 20 mA versions.

High Performance

Microprocessor-based correction provides excellent

accuracy and ambient temperature compensation.

Output Signal

1to5Vdc

Output Load

1 to 10 megohms

Supply Voltage and Current Requirements

SUPPLY VOLTAGE

9 to 15.5 V dc (Normal Operation)

SUPPLY CURRENT

3 mA

Supply Voltage Effect

The output changes less than 0.005% of calibrated

span for each 1 V change within the specified supply

voltage requirements.

Power-Up Time

Less than 5.0 seconds for output to reach first valid

measurement.

Mounting Position

The transmitter may be mounted in any orientation.

Mounting Position Effect

Any zero effect caused by the mounting position can

be eliminated by rezeroing. There is no span effect.

Field Wiring Reversal

Reversal of field wiring will not damage transmitter.

Suppressed Zero and Elevated Zero Ranges

These ranges are acceptable as long as Span and

Range limits are not exceeded. Elevated zero

applies to IGP20 and IDP10 Transmitters only.

Zero and Span Adjustments

Zero and span adjustments accomplished using the

pushbuttons on the LCD Indicator. See Figure 3

PHILIPS 958481320100 LCB
PHILIPS 958481320201 PROC PLUS
PHILIPS 958481320201 PROC+
PHILIPS 958481320400 PIF
PHILIPS 958481321200 PD216
PHILIPS 958481321210 PD212
PHILIPS 958481321220 PD208
PHILIPS 958481321300 350211080380
PHILIPS 958481321300 PSB
PHILIPS DSI020
PHILIPS OPM010
PHILIPS PR1613
PHILIPS SCM040
PHILIPS VBM010
PHILIPS VBM030
PHOENIX 2961192
PHOENIX E-UK
PHOENIX FLMC10BASE-T/FO G850